توضیحات
AnyScan SC Configuration
Multislice CT Component
Anyscan SC system combines the power of nuclear medicine with the precision of 16/8 slices fast helical CT technology.
AnyScan®’s CT module supplies precise data for the attenuation correction and provides excellent diagnostic imaging capability.
AnyScan® family offers alternative choice of 16 or 8 slice diagnostic CT or 8 slice non diagnostic CT.
The CT module consists of a high resolution ceramic detector. The rotation speed is up to 0.4* sec with a 360-degree rotation.
CT-Gantry
– Patient Aperture: 70 cm
– Scan Field of View: 50 cm (70 cm for localization and corrections)
– Acquisition Times (360°): 0.4*, 0.5, 0.7, 1, 1.5 and 2 sec.
X-Ray Generator
– Power: 60 kW for diagnostic or ≤ 10 kW for non-diagnostic CT
X-Ray Tube
– Tube Voltage: 80 – 140 kVp
– X-Ray Tube Current: 5 – 500 mA for diagnostic or 5 – 100 mA for non-diagnostic CT
– Anode Heat storage capacity: 7.5 MHU
– Automatic dose control (with ultra-low-dose & pediatric protocols)
X-Ray Detector
– Ceramic Detectors
– Number of Detector Elements: 24 x 896
– Number of DAS Rows: 16 / 8 or 8
– Detector Width (Z Dimension): 20 / 10 mm or 10 mm
Image Reconstruction
– Reconstruction Matrix: 512 x 512 or 1024 x 1024
Scan Parameters
– Acquired Slice Thickness: 16 rows of 0.625 mm / 1.25 mm or 8 rows of 1.25 mm
– Displayed Slice Thickness: 0.625 (for diagnostic CT only), 1.25, 2.5, 5, 10 mm
Dimensions, weights
Components of Anyscan SC | Length (mm) | Width (mm) | Height (mm) | Weight (kg) |
SPECT gantry | 1139 | 1714 | 1632 | 2100 |
CT gantry | 778 | 2116 | 1916 | 1350 |
Imaging Table | 2580 | 670 | 850 | 330 |
Collimators | 685 | 562 | 46-68 | 70-125 |
Collimator Carts | 750 | 560 | 1250 | 70 |
Power Distribution Unit | 1000 | 600 | 1490 | 450 |
Nucline™ Acquisition Console Workstation
Anyscan SC acquisition workstation with dual and touch screen monitor for image acquisition, settings, gantry and patient positioning.
– Intel® Core™ i7 3.0 GHz multi-core processor by Nucline™ workstation
– 16 GB RAM
– 6 TB hard disk drive
– CD-DVD-RW drive
– Keyboard, mouse
– Full DICOM 3.0 compatibility (send/ receive, print, query/retrieve)
– 24” high resolution (1920×1200) and 17” (1280×1024) touch screen LCD monitor
– Integrated Gigabit Ethernet controller
– Windows 7 64 bit Ultimate operating system (multiuser/multitasking)
– UPS power supply 3.2 KVA (optionally)
Integrated SPECT data acquisition module
– 64 independent energy channels
– Multi-channel analyzer up to 1024 channels
– 4096 x 4096 pixel image processing
– Digital corrections:
– Direct addressing TS® simulation linearity correction with FOV increasing technology
– Improved energy correction
– Uniformity correction without count rate loss
– Automatic real-time uniformity cross-correction for the different collimators
– Advanced on the fly gantry and detector alignment (COR) correction
Power Supply Requirements of Anyscan SC
SPECT-component | CT-component |
– Mains Voltage: 115/230 V (1 phase)
– Mains frequency: 45-63 Hz – Max. power: 2.4 kVA |
– Mains Voltage : 400/480 V (3 phase)
– Mains frequency: 45-63 Hz – Max. power: 65 kVA (diagnostic CT) ≤ 10 kVA (non-diagnostic CT) |
Environmental Requirements for Anyscan SC
Constant room temperature and humidity rate required for 7 days / 24 hours
– Room temperature: 20 ± 2º C (68º F ± 3.5º F)
– Max temperature change rate: 5º C/hour (7º F/hour)
– Relative humidity: 35 – 70% non-condensing
– Atmospheric pressure: 70 – 106 kPa
– Dew point < 17º C (63º F)
نقد و بررسیها
هنوز بررسیای ثبت نشده است.