Designed for facilities receiving unit doses including PET and BETA.
• Pre-programmed for 88 most commonly used radionuclides
• Large, easy-to-read backlit LCD
• Small footprint economizes workspace
• Ultra-fast response
• Automatic range selection; ranges up to 40 Curies of Tc-99m or 10 Curies of F-18
• Displays in Curies or Becquerel
• Remote Ionization Chamber
• Self-Diagnostic Software
• Desktop or wall mount display
• RS-232 bi-directional serial communications port
[/vc_column_text][/vc_column][/vc_row][vc_row][vc_column width=”1/4″][/vc_column][vc_column width=”1/2″][vc_column_text][jwp-video n=”1″][/vc_column_text][/vc_column][vc_column width=”1/4″][/vc_column][vc_column][vc_row_inner][vc_column_inner][vc_separator][/vc_column_inner][/vc_row_inner][/vc_column][/vc_row][vc_row][vc_column][vc_column_text]
Specifications of Atomlab 400 Dose Calibrator:
• Isotope Selection Keys: Ten pre-programmed – Tc-99m, Tl-201, Co-57, Cs-137, I-131, In-111, Ga-67, Xe-133, I-123, and Mo-99; seven additional keys for user-set isotopes; two new isotope keys and a full alphabetical list of 88 isotopes.
• Activity Range: 0.01 µCi to 40 Ci (.0004 MBq to 1500 GBq) of Tc-99m or 10 Ci of F-18
• Energy Range: 25 keV to 3 MeV photons
• Response Time: One to two seconds for doses greater than 200 uCi; three seconds for doses greater than 20 uCi; 50-100 seconds below 20 uCi of Tc-99m with default threshold, threshold adjustable to reduce counting time
• Detector Linearity: ±1% or 0.2 µCi, whichever is greater
• Electrometer Linearity: ±1% or 0.2 µCi, whichever is greater
• Electrometer Accuracy: ±1% or 0.2 µCi, whichever is greater
• Overall Accuracy: ±3% or 0.3 µCi, whichever is greater; overall accuracy is affected by such factors as the accuracy of the specific source calibration, geometric variations due to sample volume or configuration, detector linearity, electrometer accuracy and readout accuracy
• Repeatability: ±0.3% above 1 mCi short term (24 hr); 1% long term (one yr); exclusive of background
• Digital Calibration Dial: Four-digit LED dial display with increment/decrement keys to change the value; range is from 0.0 to 999.9
• Detector: Well-type pressurized ionization chamber, with Argon fill gas; well opening 2.75″ (7 cm), well depth 10.25″ (26 cm)
• Chamber Gas Pressure: 149KPa gauge (21.6 psig) at 20 degrees C or 250KPa absolute (36.3 psia) at 20 degrees C. IATA regulation 3.2.2.4 Exempts Gases of Division 2.2 from Dangerous Goods Regulations when transported at pressure less than 200KPa gauge (29 psig) at 20 degrees C. Device is shipped standard goods.
• Detector Shielding: .25″ (6.3 mm) lead on all sides except top well opening;
supplementary shielding available
• Chamber Bias: 355 ± 5 volts
• Environmental Operating Conditions:
– Temperature: 0-40° C
– Humidity: 0-90% rh, non-condensing
• Power Requirements: 100 to 240 VAC, 0.6 – 0.3 amps, auto switching; APS Power Supply (APS22ES-150160), for medical use.
• Line Frequency: 50/60 Hz; detachable line cord; built-in EMI filter and transient suppression
• Detector and Interface Cables: 8′ (243 cm) long, six conductor cables (two carry power, two for chassis ground, two carry serial data for digital I/O)
• Display Unit:
– Dimensions: 6.75″ w x 6″ depth x 5″ h (17.1 x 15.3 x 12.7 cm)
– Weight: 3.6 lb (1.64 kg); desktop or wall mountable
• Detector Unit:
– Dimensions: 6″ dia x 15.5″ h (15.24 x 39.37 cm)
– Well I.D.: 2.75″ dia x 10.5″ h (7 x 26.7 cm)
– Well I.D. with Liner: 2.5″ dia x 10.25″ h (6.35 x 26 cm)
– Lead Shielding: .25″ thick (6.3 mm)
– Weight: 35 lb (16 kg)
[/vc_column_text][/vc_column][/vc_row]